The WIYN One Degree Imager in 2018: An Extended 30-Detector Focal Plane
Daniel R. Harbeck, Mike Lesser, Wilson Liu +14 more
We report on the upgraded One Degree Imager (ODI) at the WIYN 3.5 meter telescope at the Kitt Peak Observatory after the focal plane was expanded by an additional seventeen detectors in spring 2015. The now thirty Orthogonal Transfer Array CCD detectors provide a total field of view of 40' x 48' on the sky. The newly added detectors underwent a design revision to mitigate reduced charge transfer efficiency under low light conditions. We discuss the performance of the focal plane and challenges in the photometric calibration of the wide field of view, helped by the addition of telescope baffles. In a parallel project, we upgraded the instrument's three filter arm mechanisms, where a degrading worm-gear mechanism was replaced by a chain drive that is operating faster and with high reliability. Three more filters, a u' band and two narrow band filters were added to the instrument's…