A linear strain seismograph*
The response of hitherto known practical forms of seismographs depends \nupon the relative motion of a pendulum and the moving ground to which its \nsupporting structure is fastened. The various kinds of seismographs differ in \nthe type of pendulum used, such as gravity, spring, torsion, and/or they differ \nin the type of magnifying and recording elements which they employ. They all \nmeasure or indicate the vibratory motion of the ground at a giver point. In contrast to these earlier forms, the strain seismograph is a nonpendular instrument. \nIt does not respond directly to vibratory movements Its operation depends \nupon variations in the distance between two points of the ground. Such \nvariations or linear strains are set up by seismic waves. A schematic drawing \nof the seismometer is shown in figure 1. A and B are two piers separated by…