A High Speed, Wide Dynamic Range Digital Data Acquisition System
Robert Auer, Bruce Weber, Eric Statler
We have previously reported on two generations of data acquisition systems for flow cytometric use which incorporated wide dynamic range active integration of the signals from the detectors and analog to digital conversion with fifteen bits of conversion resolution and twenty bits of effective dynamic range. These advances have permitted accurate computer normalization of complex dye emission spectra overlaps for multiple detectors and computer synthesis of accurate four decade logarithmic data with resolution of 256 channels per decade. We are now reporting on a third generation, higher speed data acquisition system which retains the above mentioned features of the original systems while substantially increasing the rate at which data can be acquired. The number of parameters that potentially can be acquired has also been increased from sixteen to forty eight; including pulse area…