Original Article Novel automated pulse immunoassay for human alpha-fetoprotein
Keisuke Iwata, Yukitoshi Kanayama, Tadayuki Sahara +5 more
Background: To improve current alpha-fetoprotein (AFP) assays, which are expensive and time-consuming, a specific AFP reagent has been developed for practical use in our newly developed high-speed, highly sensitive pulse immunoassay (PIA) system, in which a latex immunoagglutination reaction is carried out under a high-frequency pulse voltage, leading to an enhanced immunological reaction. Methods: We evaluated the assay performance (reproducibility, sensitivity, dilution linearity, interference) of the newly developed automated AFP PIA compared with the current AFP assay. Results: Using pooled serum samples, the within-run reproducibility resulted in a correlation variation of 3.6–4.7%. The AFP assay detection limit was determined to be 2.5 mg/L. Linear sequential dilution was found up to nearly 700 mg/L. Even up to an AFP concentration of 1.0 g/L, the prozone phenomenon was not…