arXiv.itsnick.co

an independent field guide to 23 million papers from arXiv & OpenAlex — not affiliated with arXiv.org

Physics · physics.ins-det

Qualification Tests of 474 Photomultiplier Tubes for the Inner Detector of the Double Chooz Experiment

C. Bauer, E. Borger, R. Hofacker, K. Jänner, F. Kaether, C. Langbrandtner, M. Lindner, S. Lucht, M. Reissfelder, S. Schönert, A. Stüken, C. Wiebusch

· JINST 6:P06008,2011

See on the map

Abstract

The hemispherical 10" photomultiplier tube (PMT) R7081 from Hamamatsu Photonics K.K. (HPK) is used in various experiments in particle and astroparticle physics. We describe the test and calibration of 474 PMTs for the reactor antineutrino experiment Double Chooz. The unique test setup at Max-Planck-Institut für Kernphysik Heidelberg (MPIK) allows one to calibrate 30 PMTs simultaneously and to characterize the single photo electron response, transit time spread, linear behaviour and saturation effects, photon detection efficiency and high voltage calibration.

References

The reference list for this paper has not been collected yet.

Cited by

No paper in the index cites this yet.

The cited-by list is still filling in while reference lists are collected.